Compudil 3 Specifications
| PARAMETER | |
| Accuracy and Resolution: | Full stroke of all syringes in 5000 steps. Where necessary, selected volumes are rounded down. Actual delivered volume: +/-1% of syringe volume  | 
| Speeds: | Approx. 2-15 seconds for full stroke of syringe. | 
| Precision: | CV: Better than 0.1%. | 
| Dead Volume: | Approx. 1ml, of which 95% is recoverable using the PRIME function (standard 30cm length 1.6mm I.D. reagent inlet tube, 1ml and 50ml syringes). Increased risk of detectable carry-over with dilution ratios less than 1:5, or with immiscible liquids. | 
| Chemical Resistance: | All valves, syringes and tubing are manufactured from glass or PTFE-based materials for total chemical resistance (except HF and HF compounds). | 
| Controls: | Rear panel: POWER on/off switch, socket for handset or footswitch control. RS232 socket for computer control. Front panel: 48 character display and 14 button tactile keypad.  | 
| Dimensions: | 300mm(W) x 200mm(D) x 215mm(H). | 
| Weight: | 7.5kg, (packed for shipment: 9.5kg). | 